Auto Algorithm

Obstructive Apnea
Our auto algorithm establishes a baseline of patient flow based on a moving flow signal window. The REMstar Auto detects an apnea as an 80% reduction in flow lasting at least 10 seconds.

 

Hypopnea
Our auto algorithm establishes a baseline of patient flow based on a moving flow signal window. The REMstar Auto detects a hypopnea as a 40% reduction in flow lasting at least 10 seconds, followed by a recovery breath.

Next ->